Page 40 - scales manual v1014 web
P. 40
FAIRBANKS 90-3045

GROSS/NET = Parameter select
AUTO-TARE = Data Select (enables the changing of data)

ZERO = Clears display

1. Press S2, instrument will display CXXXXX, this is the deadload counts (2000-22000)
2. Empty scale, press Zero button to set Zero Calibration
3. Press the GROSS/NET button until parameter 8 is visible.
4. Press the GROSS/NET button again and XXXXX will be displayed
5. Place weight on scale, Using AUTO-TARE, set the correct value
6. Press GROSS/NET to enter correct span reading
7. Set instrument capacity on the next step if needed
8. Press Gross/NET and the instrument will exit programming mode.
9. Remove test weights and recheck scale.

FAIRBANKS Series III

1. Remove Jumper JP3 Page 40
2. Press and hold any key and power on scale

a. “CAL” is displayed
3. Press the AUTOTARE key to select zero tracking desired

a. AO 0
b. AO 1 AZT .6d
c. AO 2 AZT 1d
d. AO 3 AZT 2d
4. Press the UNITS key to select the desired choice
5. Press AUTOTARE to display Weighing Units
a. Press UNITS to select lb or kg
6. Press AUTOTARE to display Capacity
a. Press UNITS to display desired capacity
7. Press AUTOTARE to display 12222
a. Span offset should be between 5000 and 50000 counts
8. Press the ZERO key to set the display to 0
9. Place calibration weight
10. Press AUTOTARE, scale will display a list of test weight values
11. When the appropriate weight value is displayed, press AUTOTARE
12. Install JP3