Page 38 - scales manual v1014 web
P. 38
FAIRBANKS HR2300

1. Press Pgr, scroll to CAL Menu and press ENTER
2. Enter the S code and Date or press ENTER

--S code is last two numbers of the sn and 6 digit date--
3. SC CAP-Enter scale capacity and press ENTER
4. Display shows raw counts, Press ZERO
5. LOAD- add test weights, key in value, and press ENTER
6. Scale returns to weigh mode

OR

If zero cannot be done prior to calibration use this method

1. Load scale with known weight
2. Press Pgr, scroll to “CALno0”
3. Press Enter
4. Enter the S code and date
5. Press Enter
6. Key in test weight and press Enter

Changing the Service Code on a 2300

1. Press Pgr
2. Scroll to “CAnnEd” and press Enter
3. When prompted for S code, press the keys in this order

a. Gross/Net, Units, Scroll, ID, Pgr, 4, Enter
4. Enter a 6 digit code where nUPASS is displayed
5. Press Enter

a. CLrPon is displayed, use scroll key to choose one of the two choices
i. CLrPon=password needed after power cycle
ii. CLrFin=password needed at menu entry

6. Press Enter
7. S dAtE, key in date in MMDDYY format
8. Press Enter
9. StorE
10. Gross/Net
11. Gross/Net

© 2014 ScaleResource.com Page 38