Medical Lubricated Vacuum


           3A.0  Start Up - TotalAlert Embedded Controls


           CAUTION: Vibration during shipment and              Check all voltages supplied to the LifeLine® system
           installation can loosen electrical terminals, fuse   to ensure they are the required value and phases
           inserts, and mechanical connections. Tighten as     needed by the control panel.
           necessary.
                                                               Apply power to the system and turn the disconnect
           •  Check the electrical cabinet for any broken      switches to “On”.
              components.

           •  Check that all motor starter connections are
              tight and that there are no loose objects such
              as terminal lugs, screws, nuts, etc., in the
              cabinet.

           3A.2  Initial Start-up

           CAUTION: Complete the prestart-up procedure
           before continuing with the initial start-up
           procedure

                             WARNING:

            To prevent electrical shock, ensure that
            ALL electrical power to the system is OFF,                                     Manual Override
            including the disconnect switches. The                                         Switch
            facility’s supply circuit breaker should                                       O - On Manual
            also be locked out.
                                                                                           X - Off
                                                                                           A - Automatic
           3A.2.1   Lubrication                                Figure 3A.1 Unit PCB Override Switch

           The pumps are oil lubricated.  The lubricating      Prior to actual operation, the pumps must be
           oil can be put into the pump at the oil fi ller port   checked for correct rotation.
           of the oil separator housing, observing the “MAX”
           and “MIN” position at the oil level sight glass.  (For   Inside the control cabinet, switch one of the unit
           recommended oil type, refer to Section 6.1.)  After   printed circuit boards from the manual override
           fi lling, make sure the oil fi ller port is closed.      “Off ” position to the bottom position, the default
                                                               “Automatic” mode.  Make sure the Pump Mode on
           All LifeLine® lubricated vacuum pumps are shipped   the Unit touchscreens are in the Off  position, see
           with the required amount of oil for start-up.  The   Figures 3.2 and 3.3.  (See Figure A.15 for complete
           systems are shipped with separate oil containers,   list of  Screen Toolbar Descriptions).
           which must be added before start-up.
                                                               Check for correct direction of rotation of each pump
           3A.2.2   Pump Rotation                              by pressing the “Rotation” button on touchscreen
                                                               display (found in the Service section of the Unit
           Inside the control panel, make sure that all unit   touchscreens) and observing rotation.  See Figure
           printed circuit boards are set to the manual        3.4.  The Pump Mode for each compressor must
           override “Off ” position.  This is indicated by the   be in the Off  Position for the Rotation to function.
           middle position “X” on the three-position sliding
           switch as shown in Figure 3.1.