Medical Lubricated Vacuum


           3A.0  Start Up - TotalAlert Embedded Controls



           3A.3   Initial Operation

           Start each pump by pressing “Automatic” on the
           touchscreen.  See Figure 3.5.























           Figure 3A.5 Unit Screen - Automatic Mode


                             WARNING:
               Pumps that have reached operating
               temperature may have a high surface
               temperature on the top of the exhaust
               muffl  er.

                          DO NOT TOUCH!



           Run the pump for two minutes in the correct
           rotation.  Stop the pump and top off  the oil level
           utilizing the oil fi ller port to the correct level as
           shown in the sight glass.   DO NOT OPEN THE
           FILLER PORT WHILE THE PUMP IS RUNNING.

           After testing each pump, if everything appears
           normal, put each pump into the “Automatic”
           mode and allow each pump to run until vacuum
           builds. Check for any leaks in the piping. Repair
           leaks, if needed.











           4107 9000 96.03                                 3A-4