Page 248 - Med Plaza and Cancer Center
P. 248
Medical Lubricated Vacuum
3A.0 Start Up - TotalAlert Embedded Controls
Main Screen Unit Screen # 1 Unit Screen # 2
Proximity Sensor
Power ON Light
Alarm Buzzer
Figure 3A.2 Touchscreen Controls
By observing the cooling fan of the motor, you can
determine the rotation of the pump. After pressing
the “Rotation” button on the touchscreen, there
is a 5 second delay before the pump will start for
a brief amount of time. Pump rotation should be
counterclockwise when looking at the rear of the
motor. Directional arrows are located on each
pump.
If the pumps are rotating in the wrong direction,
rotation can be reversed by switching any two
main power leads to the panel. Correct rotation
should be confi rmed in the previous manner.
Figure 3A.3 Unit Screen - Off Position
WARNING:
Do not allow the vacuum pump to
run backwards.
Repeat the process of switching the Unit printed
circuit boards from the manual override “Off ”
position to the default “Automatic” position and
testing rotation.
Figure 3A.4 Unit Screen - Service: Rotation