Medical Lubricated Vacuum


            3A.0  Start Up - TotalAlert Embedded Controls



                                Main Screen                Unit Screen # 1                  Unit Screen # 2





               Proximity Sensor



                Power ON Light


                  Alarm Buzzer







            Figure 3A.2 Touchscreen Controls


                                                                 By observing the cooling fan of the motor, you can
                                                                 determine the rotation of the pump. After pressing
                                                                 the “Rotation” button on the touchscreen, there
                                                                 is a 5 second delay before the pump will start for
                                                                 a brief amount of time. Pump rotation should be
                                                                 counterclockwise when looking at the rear of the
                                                                 motor.  Directional arrows are located on each
                                                                 pump.

                                                                 If the pumps are rotating in the wrong direction,
                                                                 rotation can be reversed by switching any two
                                                                 main power leads to the panel.  Correct rotation
                                                                 should be confi rmed in the previous manner.

            Figure 3A.3 Unit Screen - Off  Position
                                                                                   WARNING:
                                                                      Do not allow the vacuum pump to
                                                                                 run backwards.



                                                                 Repeat the process of switching the Unit printed
                                                                 circuit boards from the manual override “Off ”
                                                                 position to the default “Automatic” position and
                                                                 testing rotation.









            Figure 3A.4 Unit Screen - Service: Rotation

                                                             3A-3                                 4107 9000 96.03