Wiring Schematic 6: General Fault / Auxiliary Relays
Auxiliary Relay General Fault Relay
NOTE:
N/O (pin 6) N/C (pin 5) General Fault Relay activates when ANY
alarm on panel is in progress. Pressing
Common (pin 7) Common (pin 4)
MUTE on Annunciator Module DOES
N/C (pin 8) N/O (pin 3)
NOT reset relay. Relay deactivates when
alarm condition is corrected.
Auxiliary Relay (optional) activates when
ANY audible alarm on panel is in progress.
ribbon to Pressing MUTE on Annunciator Module
adjacent WILL deactivate relay until audible alarm is
module again reactivated.
Relay contact ratings are 2A @ 30 VDC/0.5
A @ 125 VAC.
Annunciator Module
Wiring Schematic 7: Digital Display Module High / Low Relays
High Alarm Relay Low Alarm Relay NOTE:
High Alarm Relay activates when high line
Common (pin 9) N/C (pin 8) pressure alarm on digital display module is
in progress. When relay activates normally
N/C (pin 10) Common (pin 7)
closed contacts open. Pressing MUTE
on Annunciator Module DOES NOT deacti-
vate relay. Relay deactivates when high line
pressure alarm condition is corrected.
ribbon to
Low Alarm Relay activates when low line
adjacent
pressure alarm on digital display module is
module
in progress. When relay activates normally
closed contacts open. Pressing MUTE
on Annunciator Module DOES NOT deacti-
vate relay. Relay deactivates when low line
pressure alarm condition is corrected.
Digital Display Module
Relay contact ratings are 2A @ 30 VDC/0.5
A @ 125 VAC.
19
Part No. 6-847684-00 Rev. E00