Typical Master/Area Alarm Wiring Examples

                                                                       MASTER ALARM PANEL #1
                                                       pin 6 (-)   pin 5 (+)




                      Oxygen
                      Sensor
                      Module
                  Connected to main                                                             MAIN LINE
                   oxygen pipeline
                                                       Digital Display Module






                                                   pin 6 (com)          pin 1 (+)
                                                                                                   O2 Secondary
                                                                                                     Supply
                Com
                                                                                                    O2 Reserve
                                                                                                     In Use
                Secondary Supply -
                                                                                                    O2 Reserve
              from liquid O2 manifold
                                                                                                      Low
                 Com

                                                        Multi-Signal Module
                 Reserve In Use -
               from liquid O2 system


                Com
                                                    pin 6 (com)         pin 1 (+)                  O2 Secondary
                                                                                                     Supply
                    Reserve Low -
                                                                                                    O2 Reserve
               from liquid O2 HP back-up
                                                                                                     In Use
                                                                                                   O2 Reserve
                                                                                                      Low
                       NOTE:
                       Switch contacts must be dry,
                       normally-closed type.  Alarm
                       activates when contacts open.    Multi-Signal Module





                                                      pin 6 (-)     pin 5 (+)



                      Oxygen
                      Sensor
                      Module
                  Connected to main
                                                                                                MAIN LINE
                   oxygen pipeline
                                                       Digital Display Module

                                                                         MASTER ALARM PANEL #2

                Example 2: Two master alarm panels with digital displays -
                            liquid manifold/high-pressure reserve (oxygen shown).
                                                                                                                 23
                                                                                            Part No. 6-847684-00 Rev. E00