Typical Master/Area Alarm Wiring Examples

                                                                    MASTER ALARM PANEL #1
                                                   pin 6 (-)    pin 5 (+)




                   Oxygen
                   Sensor
                   Module

              Connected to main                                                              MAIN LINE
                oxygen pipeline
                                                    Digital Display Module





                                                                     pin 1
                                                                                               O2 Main Tank
                                                                                                Liquid Low
             Com                                                                                O2 Reserve
                                                                                                  In Use
           Main Tank Liquid Low -                                                               O2 Reserve
            from main bulk tank                                                                 Liquid Low
                                                                                              O2 Reserve Tank
            Com                                                                                Pressure Low

             Reserve In Use -
                                                     Multi-Signal Module
            from bulk O2 system
             Com
                                                                     pin 1
          Res. Tank Liquid Low -
            from main bulk tank                                                                O2 Main Tank
                                                                                              Liquid Level Low
            Com                                                                                 O2 Reserve
                                                                                                  In Use
                                                                                                O2 Reserve
         Res. Tank Pressure Low -
                                                                                              Liquid Level Low
           from reserve bulk tank
                                                                                              O2 Reserve Tank
                                                                                               Pressure Low
                  NOTE:
                  Switch contacts must be dry,
                                                     Multi-Signal Module
                  normally-closed type.  Alarm
                  activates when contacts open.


                                                   pin 6 (-)    pin 5 (+)


                  Oxygen
                   Sensor
                  Module

              Connected to main
                                                                                             MAIN LINE
                oxygen pipeline
                                                    Digital Display Module
                                                                   MASTER ALARM PANEL #2


                                                   Example 3: Two master alarm panels with digital displays -
                                                               liquid bulk/liquid reserve (oxygen shown).
         24
              Part No. 6-847684-00 Rev. E00