Configuring and Browsing the Master Alarm by Using the Web Pages
1
Chapter 1: Introduction
b. Without the NetBios Name Service enabled, the device must be
accessed by using its IP address.
c. To learn the Master Alarm’s IP address, press the Test key on its front
panel.
4. Click on the Location pull down menu arrow. The available location
options display on the screen (see figure 1-35 on page 1-85).
Figure 1-35. Setup Device—Location
The options are as follows:
a. PBX—Public broadcast system room
b. OR—Operating room
c. ER—Emergency room
d. CCR—Critical care unit
e. PACU—Post anesthesia care unit
f. ENG—Engineering room
g. Nurse Station—Nurses’ station
h. Special—Special care unit
i. General—General medical unit
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 85