Page 717 - Med Plaza and Cancer Center
P. 717

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   Configuring the Area Alarm by Using the Front Panel
                   Chapter 1: Introduction



                                 WARNING:
                                 In operating systems, variations in pressure can cause disruption of
                                 vital life support equipment. Do not vary pressures without the
                                 appropriate medical facility personnel’s permission and knowledge.
                                 Varying pressure can endanger patient welfare.

                                 3. Before changing any pressure setpoints, notify the appropriate medical
                                     facility personnel.

                                 4. Configure   A hi   . This is the highest that the pressure can get without
                                     alarming.

                                 The Area Alarm Modules are factory-programmed to NFPA 99 standard values
                                 and with the units of measure, either metric or English, that were specified by
                                 the customer (see table 1-4 on page 1-30).


                                 5. Configure   A Lo   . This is the lowest that the pressure can get without
                                     alarming.

                                                      . This is the amount of time that the alarm stays silent.
                                 6. Configure A rt
                                     The available options are as follows:

                                      •   No , audible alarm does not repeat.
                                      •   30 , audible alarm repeats every 30 minutes.


                                      •   60 , audible alarm repeats every 60 minutes.
                                      •   90 , audible alarm repeats every 90 minutes.


                                      •   120   , audible alarm repeats every 120 minutes.

                                 NOTE:
                                 To comply with the requirements of the City of New York Department of
                                 Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
                                 the alarm return time to “never.” Set the alarm to return every 30 minutes.





















                   Page 1 - 120          MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)