PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03
Configuring the Area Alarm by Using the Front Panel
1
Chapter 1: Introduction
Returning to Normal Operating Mode
1. To return to normal operation and save any changes that you have made, do
one of the following:
• Press the configuration button (CE) or (CF) (see figure 1-57 on page
1-117).
• Turn the power supply switch (DR) to the OFF position, and then
back to the ON position (see figure 1-56 on page 1-115).
If you do not press any keys for 10 minutes, the alarm returns to normal
operation and saves any changes that you have made.
2. While GASt is being displayed, you can view the type of gas that the
Area Alarm Module is monitoring by pressing the Increment key or the
Decrement key. For a list of available gases, refer to table 1-31 on page 1-
119.
NOTE:
The type of gas or vacuum being monitored cannot be changed.
Table 1-31. Area Alarm Module Gas Types
Display Shows Gas type
Oxygen
oy9n
Medical air
M7Ed
Vacuum
UAcu
WAGD (waste anesthetic gas disposal)
WJA9
Nitrous oxide
nito
Nitrogen
nit9
Carbon dioxide
cArb
Helium
hELi
Oxygen and carbon dioxide mixture
Oycd
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 119