“Oil-Less” Scroll Medical Air


            5.0  Maintenance



            5.1  Maintenance Schedule                            Release all pressure from the package
                                                                 before removing, loosening, or servicing any
                                                                 covers, guards, fi ttings, connections, or other
                              WARNING:
                                                                 devices.
             BEFORE STARTING ANY MAINTENANCE
             PROCEDURES, DISCONNECT ALL POWER                    Never perform any maintenance functions
             TO THE PACKAGE.                                     while the unit is in operation.




             Maintenance Schedule
             Item                            Frequency                   Action
             Check condensate in tank        Daily                       Open manual drain valve, check auto drain

             Check operation of safety valve  Weekly                     Manually release pressure

             Check inlet air fi lter(s)       Weekly                      Inspect and clean or replace

             Check nuts, bolts, fi ttings, etc.  Monthly                  Inspect and tighten

             Check belt tension              Monthly                     Inspect and tighten or replace

             Check fl ow through orifi ce of dew   Every 6 months          Check for fl ow blockage
             point sensor

                                                                         Purchase 6-Month Service Kit
             Calibrate CO transmitter        Every 6 months
                                                                         See Appendix B

                                                                         Verify dew point sensor accuracy
             Check dew point sensor accuracy  Yearly
                                                                         (contact BeaconMedæs)
             Replace compressor inlet fi lters
                                             Yearly                      Purchase 1-Year System Basic Service Kit
             and dryer pre-fi lters & afterfi lters
                                                                         Purchase 2-Year System Sensor Kits
             Replace DP and CO transmitters  Every 2 years
                                                                         See Appendix B and C

                                                                         Purchase 3-Year Dryer Extended Service Kit
             Replace dryer desiccant         Every 3 years
                                                                         See Appendix A
             Lubricate motor                 Yearly                      See Section 6.3
             Orbiting & drive bearing        10,000 hours (2, 3, 5, 10 Hp)   Contact BeaconMedæs Technical Support
             lubrication*                    8,000 hours (7.5, 15 Hp)     department
                                             10,000 hours (2, 3, 5, 10 Hp)   Contact BeaconMedæs Technical Support
             Replace compressor tip seals*
                                             8,000 hours (7.5, 15 Hp)    department

            *An authorized BeaconMedæs Service Technician should perform compressor lubrication and tip seals
            replacement.  Please contact BeaconMedæs Customer Service department at 1-800-756-2590 prior to
            10,000 operating hours (8,000 hours for 7.5, 15 Hp) to schedule this maintenance.



                                                             5-1                                  4107 9000 69.06