“Oil-Less” Scroll Medical Air
5.0 Maintenance
5.1 Maintenance Schedule Release all pressure from the package
before removing, loosening, or servicing any
covers, guards, fi ttings, connections, or other
WARNING:
devices.
BEFORE STARTING ANY MAINTENANCE
PROCEDURES, DISCONNECT ALL POWER Never perform any maintenance functions
TO THE PACKAGE. while the unit is in operation.
Maintenance Schedule
Item Frequency Action
Check condensate in tank Daily Open manual drain valve, check auto drain
Check operation of safety valve Weekly Manually release pressure
Check inlet air fi lter(s) Weekly Inspect and clean or replace
Check nuts, bolts, fi ttings, etc. Monthly Inspect and tighten
Check belt tension Monthly Inspect and tighten or replace
Check fl ow through orifi ce of dew Every 6 months Check for fl ow blockage
point sensor
Purchase 6-Month Service Kit
Calibrate CO transmitter Every 6 months
See Appendix B
Verify dew point sensor accuracy
Check dew point sensor accuracy Yearly
(contact BeaconMedæs)
Replace compressor inlet fi lters
Yearly Purchase 1-Year System Basic Service Kit
and dryer pre-fi lters & afterfi lters
Purchase 2-Year System Sensor Kits
Replace DP and CO transmitters Every 2 years
See Appendix B and C
Purchase 3-Year Dryer Extended Service Kit
Replace dryer desiccant Every 3 years
See Appendix A
Lubricate motor Yearly See Section 6.3
Orbiting & drive bearing 10,000 hours (2, 3, 5, 10 Hp) Contact BeaconMedæs Technical Support
lubrication* 8,000 hours (7.5, 15 Hp) department
10,000 hours (2, 3, 5, 10 Hp) Contact BeaconMedæs Technical Support
Replace compressor tip seals*
8,000 hours (7.5, 15 Hp) department
*An authorized BeaconMedæs Service Technician should perform compressor lubrication and tip seals
replacement. Please contact BeaconMedæs Customer Service department at 1-800-756-2590 prior to
10,000 operating hours (8,000 hours for 7.5, 15 Hp) to schedule this maintenance.