“Oil-Less” Scroll Medical Air


            4.0  Trouble Shooting



             Problem                    Possible Causes                   Solution

             Failure to start           Main power disconnected           Turn on main power

                                                                          Change power supply phase on incoming
                                                                          power

                                        Power failure                     Restore power

                                        Main fuse blown                   Replace fuse

                                        Fuse blown in control circuit     Replace fuse

                                        Overload tripped on starter       Reset & check for system overload

                                                                          Allow unit to cool; reset alarm & check for
                                        High temperature sensor activated
                                                                          over temperature condition

                                        Pressure sensor open              Adjust or replace sensor
                                        Loose or faulty connection        Check & tighten all wire connections

             Power failure              Main fuse blown                   Replace fuse

                                        Fuse blown in control circuit     Replace fuse

             Compressor shuts off       Overload tripped on starter       Reset & check for system overload
             unexpectedly
                                        Pressure sensor failure           Replace

                                        High inlet vacuum switch activated  Check for dirty/clogged inlet fi lter

                                                                          Check for inlet piping restriction

                                        High temperature sensor activated  Allow unit to cool; reset alarm & check for
                                                                          over temperature condition

             High temperature alarm                                       Allow unit to cool; reset alarm & check for
                                        High temperature sensor activated
                                                                          over temperature condition
             Motor overheating          Low voltage                       Check for proper supply voltage

                                        V-belt too tight                  Adjust belt tension

                                        Defective motor                   Contact BeaconMedæs

             High inlet vacuum alarm    Inlet piping restriction          Check inlet piping

                                        Clogged inlet fi lter              Replace inlet fi lter