Page 191 - Womens Pavilion
P. 191

“Oil-Less” Scroll Medical Air


           3.0  System Operation



           section of the Unit touchscreens - See Section
           D.3.7 for additional information) and observing
           rotation.  See Figure 3.4.  The Pump Mode for each
           compressor must be in the Off  Position for the
           Rotation to function.


                             WARNING:

               DO NOT RUN THE COMPRESSOR
               BACKWARDS!






                                                               Figure 3.5 Unit Screen - Automatic Mode

                                                               After testing each compressor, if everything
                                                               appears normal, put each compressor into the
                                                               “Automatic” mode and allow each compressor to
                                                               run until pressure builds in the air receiver. The lag
                                                               compressor should stop fi rst when the pressure
                                                               reaches its set point on the controller. Pushing the
                                                               reset button on the control panel can now reset
                                                               the lag alarm. See Figure 3.6.  The lead compressor
                                                               should stop when the pressure reaches 110 psig.
           Figure 3.4 Unit Screen - Service: Rotation          Check for any leaks in the piping up to the inlet
                                                               isolation valves of the dryers. Repair leaks, if
           Rotation direction arrows are located on the        needed.
           belt guard (rotation is counter clockwise, facing
           the compressor pulley). Correct the rotation, if
           required, by switching the motor leads at the
           starter.

           REMOVE POWER BEFORE WORKING ON ANY
           ELECTRICAL CONNECTIONS.


           Repeat the process of switching the Unit printed
           circuit boards from the manual override “Off ”
           position to the default “Automatic” position and
           testing rotation.


           Start each compressor by pressing “Automatic”
           on the touchscreen. See Figure 3.5.  Allow each
           compressor module to operate for a short time (15                              Reset Button
           to 30 seconds) and check for any unusual noises     Figure 3.6 Main Screen - Reset Button
           or vibrations.  Switch the compressor back to the
           “Off ” position on the touchscreen.


           4107 9000 69.06                                  3-4
   186   187   188   189   190   191   192   193   194   195   196