“Oil-Less” Scroll Medical Air


            3.0  System Operation



            When the dryer is in the “Automatic” mode, the       3.6  Emergency Shutdown / Alarms
            dew point monitor controls the dryer purge, and
            purging depends on the dew point condition.          The following conditions may arise during
            See Figure 3.7. When the dew point reading is        operation.
            above the setpoint of -10°C (14˚F), the dryer will
            function normally (one tower at system presure,
            one tower at 0 psig). When the dew point is below
            the setpoint of -10°C (14˚F), the purge valve
            will close. In this condition, both towers will be
            approximately 100 psig (dependent upon tank
            pressure) and the dryer will not shift towers until
            the dew point is above -10°C (14˚F).








                                                                 Figure 3.8 Unit Screens - Shutdown and Alarms

                                                                 Motor Overload Shutdown - This will shut down
                                                                 the compressor in question and will not re-start
                                                                 until the reset button on the starter inside the main
                                                                 control cabinet is reset and the appropriate reset
                                                                 button is pressed on the control panel display.
                                                                 See “Motor overheating” in the Troubleshooting
                                                                 Section 4.0.
            Dryer 2 is in Automatic mode, Dryer 1 is in Off  position
                                                                 High Air Temperature Shutdown - This will shut
            Figure 3.7 Main Screen - Dryer Operation             down the compressor in question and will not re-
                                                                 start until the appropriate button is pressed on the
            3.5  Normal Shutdown                                 control panel. Before allowing the unit to re-start,
                                                                 the condition should be checked (see “Compressor
            Dryer in Automatic Mode                              shuts off  unexpectedly” in the Troubleshooting
                                                                 Section 4.0). Even after resetting the alarm and
            Disconnect switches-OFF                              putting the compressor in “Automatic” mode,
                                                                 the unit may not re-start, depending on system
            Main power source-OFF                                sequencing and system pressure.

            Hospital shutoff  valve-CLOSED                       High Inlet Vacuum Shutdown - This will shut
                                                                 down the compressor in question and will not re-
            Air receiver manual tank drain-OPEN                  start until the appropriate button is pressed on the
                                                                 control panel. Before allowing the unit to re-start,
            Pressure gauge decreasing to 0 psi (0 kPa)           the condition should be checked (see “Compressor
                                                                 shuts off  unexpectedly” in the Troubleshooting
            Close air receiver manual tank drain when pressure   Section 4.0). Even after resetting the alarm and
            decreases to 0 psi (0 kPa)                           putting the compressor in “Automatic” mode,
                                                                 the unit may not re-start, depending on system
                                                                 sequencing and system pressure.