Page 195 - Womens Pavilion
P. 195
“Oil-Less” Scroll Medical Air
3.0 System Operation
High Air Temperature Alarm - This will not shut Lag Unit Running Alarm - This alarm will activate
down the compressor in question but instead is if the last available compressor unit comes on.
a warning that a shutdown is likely to occur. The (See Section 3.4 for normal operation) To silence
condition should be checked immediately (see the alarm, press the horn silence button. In the
“High temperature alarm” in the Troubleshooting event the lag alarm is persistent, check to see if any
Section 4.0) to avoid a compressor shutdown. leaks or valves are open downstream or reduce the
system load.
Control Circuit Alarm - This will not shut down
the compressor in question but instead is a Ambient Temperature Alarm - This alarm will
notifi cation that there is a loss of communication activate when the temperature in the room
between printed circuit boards within the control exceeds the set point. The audible alarm will not
panel. See Appendix D on Control System for sound but a touchscreen will show an active alarm
troubleshooting. and record it in the event log. The alarm remains
latched until the alarm condition is reset by the
operator.
Dryer Tower Switching Failure Alarm (Optional)
For units with this feature, there is one pressure
switch that senses pressure inside each dryer
tower for a total of two pressure switches for each
dryer. If the pressure in both towers of the same
dryer fall below 50 psig, the “Dryer # Switch” on
the alarm screen will change from Green to Red
and the remote dryer fault contacts (one for each
dryer) will change condition (refer to the wiring
diagram that came with the unit for terminal
numbers). A short time delay (1.5 seconds) is
started when both pressure switches fall below 50
Figure 3.9 Main Screen - Alarms psig to eliminate false alarms during normal tower
switching. To eliminate alarms during periods
Dew Point Alarm - This alarm will activate if the when the dryer is turned “off ” for maintenance
dew point exceeds the alarm setting. To silence the or other reasons, the dryer tower switching failure
alarm, press the horn silence button. If the situation alarm is bypassed for that dryer.
does not correct itself through normal dryer use,
see Appendix B, section B.7 Troubleshooting for Low Dryer Outlet Pressure Alarm (Optional)
possible causes and solutions. The alarm remains For units with this feature, there is an adjustable
latched until the alarm condition is reset by the pressure switch located upstream of the dryer
operator. check valve that will alarm if the pressure in that
line drops below the 75 psig factory setpoint. In
CO Level Alarm - This alarm will activate if the the event of an alarm, the “Dryer # Pressure” on
CO level exceeds the 10 ppm set point. To silence the alarm screen will change from Green to Red
the alarm, press the horn silence button. See and a remote set of contacts will change condition
Appendix C, section C.10 Troubleshooting for (refer to the wiring diagram that came with the
possible causes and solutions. The alarm remains unit for terminal numbers).
latched until the alarm condition is reset by the
operator.
4107 9000 69.06 3-8