Page 195 - Womens Pavilion
P. 195

“Oil-Less” Scroll Medical Air


           3.0  System Operation



           High Air Temperature Alarm - This will not shut     Lag Unit Running Alarm - This alarm will activate
           down the compressor in question but instead is      if the last available compressor unit comes on.
           a warning that a shutdown is likely to occur.  The   (See Section 3.4 for normal operation)  To silence
           condition should be checked immediately (see        the alarm, press the horn silence button.  In the
           “High temperature alarm” in the Troubleshooting     event the lag alarm is persistent, check to see if any
           Section 4.0) to avoid a compressor shutdown.        leaks or valves are open downstream or reduce the
                                                               system load.
           Control Circuit Alarm - This will not shut down
           the compressor in question but instead is a         Ambient Temperature Alarm -  This alarm will
           notifi cation that there is a loss of communication   activate when the temperature in the room
           between printed circuit boards within the control   exceeds the set point.  The audible alarm will not
           panel.  See Appendix D on Control System for        sound but a touchscreen will show an active alarm
           troubleshooting.                                    and record it in the event log.   The alarm remains
                                                               latched until the alarm condition is reset by the
                                                               operator.

                                                               Dryer Tower Switching Failure Alarm (Optional)
                                                               For units with this feature, there is one pressure
                                                               switch that senses pressure inside each dryer
                                                               tower for a total of two pressure switches for each
                                                               dryer. If the pressure in both towers of the same
                                                               dryer fall below 50 psig, the “Dryer # Switch” on
                                                               the alarm screen will change from Green to Red
                                                               and the remote dryer fault contacts (one for each
                                                               dryer) will change condition (refer to the wiring
                                                               diagram that came with the unit for terminal
                                                               numbers).  A short time delay (1.5 seconds) is
                                                               started when both pressure switches fall below 50
           Figure 3.9 Main Screen - Alarms                     psig to eliminate false alarms during normal tower
                                                               switching.  To eliminate alarms during periods
           Dew Point Alarm - This alarm will activate if the   when the dryer is turned “off ” for maintenance
           dew point exceeds the alarm setting.  To silence the   or other reasons, the dryer tower switching failure
           alarm, press the horn silence button.  If the situation   alarm is bypassed for that dryer.
           does not correct itself through normal dryer use,
           see Appendix B, section B.7 Troubleshooting for     Low Dryer Outlet Pressure Alarm (Optional)
           possible causes and solutions. The alarm remains    For units with this feature, there is an adjustable
           latched until the alarm condition is reset by the   pressure switch located upstream of the dryer
           operator.                                           check valve that will alarm if the pressure in that
                                                               line drops below the 75 psig factory setpoint.  In
           CO Level Alarm -  This alarm will activate if the   the event of an alarm, the “Dryer # Pressure” on
           CO level exceeds the 10 ppm set point.  To silence   the alarm screen will change from Green to Red
           the alarm, press the horn silence button.  See      and a remote set of contacts will change condition
           Appendix C, section C.10 Troubleshooting for        (refer to the wiring diagram that came with the
           possible causes and solutions. The alarm remains    unit for terminal numbers).
           latched until the alarm condition is reset by the
           operator.




           4107 9000 69.06                                  3-8