Page 496 - Womens Pavilion
P. 496

TotalAlert Infinity™ NFPA Medical Gas Notification System



            •  Alarm Type Tab                                    •  Initialize Modules Tab

            The ALARM TYPE screen defines which type             The PRIMARY SETTINGS screen will now dis-
            of alarm will be displayed and drives later          play with ALARM TYPE checked and INITIAL-
            setup functions.                                     IZE MODULES active.

                                                                 •  Touch the INITIALIZE MODULES tab.

















                      Figure 125: Alarm Type Tab
            •  Touch the ALARM TYPE tab.
            •  Choose a code style based on code re-                   Figure 127: Initialize Modules Screen
               quirements.
            •  Touch the COMBO tab.                              The INITIALIZE MODULES screen is used to set
                                                                 up the boards that are installed. The boards
                                                                 function as the input/output hardware of the
                                                                 alarm.

                                                       Save      At this point, you have two options:
                                                       Erase     1.  Either touch the Scan button to prompt
                                                                    the alarm to autofill the screen for you.

                                                                 OR
                                                                 2.  Complete the screen by selecting the
                                                                    boards installed in the alarm.
              Figure 126: Choosing Code and Alarm Styles
                                                                 Touch the Save button when you see checks
            •  Once the desired tabs are selected, touch         beside your selections.
                the Save button to save the selection and        Notes:
                go back to the PRIMARY SETTINGS screen.
                                                                 1.  Select the NFPA/CSA SIGNAL / B1x board
            Note: To clear your selections, touch the               during the configuration of a combination
            Erase button.                                           alarm.
                                                                 2.  Select Gas Input / B60 or Gas Combina-
                                                                    tion / B65 depending on hardware in-
                                                                    stalled in the alarm.








                                                             4-29                               4107 9016 58.02