TotalAlert Infinity™ NFPA Medical Gas Notification System
Gas Sensors Tab To select the first gas, touch the D01=NONE/
The PRIMARY SETTINGS screen will now dis- BLANK tab.
play with ALARM TYPE and INITIALIZE MOD- • Touch the right and left arrows at the
ULES checked, GAS SENSORS active, and the bottom of the screen to scroll through
last tab grayed out. additional gas choices. Select the gas that
• Touch the GAS SENSORS tab is wired into the D01 input on the GAS
INPUT / B60 board.
• Touch the Save button when you see a
check beside your selection. (If an alarm
sounds, press the Mute button on the
panel under the screen and continue with
setup.)
• Repeat to add additional gases.
• Touch the X button in the upper right to
close the GAS SENSORS screen and return
to the PRIMARY SETTINGS screen. The
Figure 128: B60 Gas Input Screen
alarm will retain your information.
B60 Board B65 Board
Users may select up to six different gases on The GAS SENSORS screen defaults to the B60
the GAS INPUT screen. These gases correspond board, so users will see a MODULE NOT INI-
to the pressure sensors wired into the GAS TIALIZED message.
INPUT / B60 board or the Gas Combination /
B65 board.
Note: This setup accommodates:
• One master badge (with 10 source badges)
and two area badges = three total badges
• One master badge (with 10 source badges)
and four area badges = five total badges
• One master badge (with 10 source badges)
and six area badges = seven total badges
Figure 130: B60 Module not initialized
• Touch the B65 tab to go to the GAS COM-
BINATION screen.
Figure 129: Gas Choices