TotalAlert Infinity™ NFPA Medical Gas Notification System



           Gas Sensors Tab                                     To select the first gas, touch the D01=NONE/

           The PRIMARY SETTINGS screen will now dis-           BLANK tab.
           play with ALARM TYPE and INITIALIZE MOD-            •  Touch the right and left arrows at the
           ULES checked, GAS SENSORS active, and the               bottom of the screen to scroll through
           last tab grayed out.                                    additional gas choices. Select the gas that
           •  Touch the GAS SENSORS tab                            is wired into the D01 input on the GAS
                                                                   INPUT / B60 board.
                                                               •  Touch the Save button when you see a
                                                                   check beside your selection. (If an alarm
                                                                   sounds, press the Mute button on the
                                                                   panel under the screen and continue with
                                                                   setup.)

                                                               •  Repeat to add additional gases.
                                                               •  Touch the X button in the upper right to
                                                                   close the GAS SENSORS screen and return
                                                                   to the PRIMARY SETTINGS screen. The
                  Figure 128: B60 Gas Input Screen
                                                                   alarm will retain your information.
           B60 Board                                           B65 Board
           Users may select up to six different gases on       The GAS SENSORS screen defaults to the B60
           the GAS INPUT screen. These gases correspond  board, so users will see a MODULE NOT INI-
           to the pressure sensors wired into the GAS          TIALIZED message.
           INPUT / B60 board or the Gas Combination /
           B65 board.

           Note: This setup accommodates:
           •  One master badge (with 10 source badges)
              and two area badges = three total badges
           •  One master badge (with 10 source badges)
              and four area badges = five total badges

           •  One master badge (with 10 source badges)
              and six area badges = seven total badges


                                                                    Figure 130: B60 Module not initialized

                                                               •  Touch the B65 tab to go to the GAS COM-
                                                                   BINATION screen.












                       Figure 129: Gas Choices