TotalAlert Infinity™ NFPA Medical Gas Notification System
To select the first gas, touch the D01=NONE/
BLANK tab.
Figure 131: B65 Board
• Users may select up to three different
gases on the GAS COMBINATION screen. Figure 133: Gas Choices
These gases correspond to the pressure • Use the right and left arrows at the bot-
sensors wired into the GAS COMBINATION tom of the screen to scroll through ad-
/ B65 board. ditional gas choices. Select the gas that
Note: This setup accommodates: is wired into the D01 input on the GAS
• One master badge (with 10 source badg- COMBINATION / B65 board.
es)and two area badges = three total • Touch the Save button when you see a
badges check beside your selection. (If an alarm
• One master badge (with 10 source badg- sounds, press the Mute button on the
es) and four area badges (only three can panel under the screen and continue with
be configured) = five total badges setup.)
• One master badge (with 10 source badg- • The GAS COMBINATION screen will now
es) and six area badges (only three can be display with the D01=[X] checked
configured) = seven total badges
• Repeat the above steps to select up to two
additional gases.
NOTE:
See the end of this section for information
on the Miscellaneous Tab. See section 4.6
for information on the Active Alarm Opt
Tab.