TotalAlert Infinity™ NFPA Medical Gas Notification System
A = [Gas Input] Tab C = [Gas Name] Tab
The B50 should be green and there should be Use the keyboard to assign a device name.
4 choices (I01, I02, I03, I04). Select the wired Type in the name of the device to be moni-
input to be displayed on that badge. tored. Touch the [<ÁÑ] to toggle to symbols
Touch the tab for the appropriate wired gas and extended Latin Characters. Touch the
sensor. [1AZ] key to toggle back to letters and num-
bers.
Touch the Save button when complete.
Figure 148: B50 Tab, 4-20 mA
Touch the Save button to close the screen. Figure 150: Gas Text Keyboard
B = Gas Desc Tab
To assign a color code for the device: D = Area Monitored Tab
To add locations to be monitored:
1. Scroll through the codes using the right
and left arrows at the bottom of the 1. Touch an unused location tab.
screen. Custom colors are located on 2. Using the keyboard, enter the name of the
pages 7-8. location that you would like to monitor
2. Touch the tab for the appropriate code. (e.g., ROOM 1). Enter multiple locations
3. Touch the Save button when you see a as needed (e.g., Room 1, Room 2, Room
check beside your selection. 5).
3. A check appears when the location is se-
lected, the save button has to be pressed
if the selection is new.
To edit the location information, touch the
Pencil button.
Figure 149: Gas Color Code Tab
4107 9016 58.02 4-36