TotalAlert Infinity™ NFPA Medical Gas Notification System



           Active Alarm                                        The B40 (or B41, B42, B43) board must be

           To setup active alarm General Message:              initialized.  Touch Primary Settings from the
           1.  Touch the General Message tab to activate       right side menu, then select “Initialize Mod-
              general message.                                 ules”. Press the “Scan” button to make sure
           2.  Touch the gray text box at bottom of screen     the B40 board is set up, as it should turn
              to type in general message.                      green. Once the B40 board is green, press the
           3.  Using the keyboard, enter the message that      “Save” icon to exit the screen.  Exit the “Pri-
              you would like to be active when in alarm        mary Settings” screen.
              mode and save.

           To setup active alarm Option Message:
           1.  Touch the Optional Message tab to activate
              optional message.
           2.  Touch the 01 tab.
           3.  Using the keyboard, enter the optional mes-
              sage that you would like to be active when
              in alarm mode and save.


           Configuring Additional 4-20mA Devices
           Configure additional 4-20mA devices by touch-             Figure 155: Initialize Modules Screen
           ing the number on the numeric display that          The next step is to select the “Relay Outputs”
           corresponds with the gas badge and repeating        from the menu on the Configuration Screen.
           the steps listed above.                             The B40 board should be highlighted in
                                                               Green (or other Relay Output board installed
           4.7.2 Relay Output Board (B4X                       in the rough-in box).
           Board)                                              Select the signal that corresponds to the
           To configure a Relay Output,  start on the          output wires on the relay board (Y01 for
           CONFIGURATION screen. Users are always              example).
           prompted to enter the password when ac-
           cessing this screen.
           Type the password, and touch the Enter but-
           ton.

           The CONFIGURATION screen displays.









                                                                       Figure 156: Relay Outputs Screen

                                                               After selecting the signal, you will need to
                                                               assign it to an alarm signal to send.  Select
                                                               “Wired” on the Relay Output screen if the
                                                               signal to be relayed is wired into the rough-in
                  Figure 154: Configuration Screen             box.

           4107 9016 58.02                                 4-38