TotalAlert Infinity™ NFPA Medical Gas Notification System
Wired Ethernet Settings
On the Miscellaneous screen, touch the
WIRED ETHERNET tab (Figure 162).
Figure 161: Gas Sensor Alarm Fault
4.7.3 Wired Ethernet Setup
Figure 162: Miscellaneous Screen
The TotaAlert Infinity™ Notification System The WIRED ETHERNET screen allows for
allows for networking via standard Ethernet. customization of the Ethernet settings. All
Any alarm panel can be remotely observed of the tabs on this screen are set to default
on the network using a standard web brows- status.
er.
IP Address using DHCP method: Upon power-
up of the system, the device will search for
a DHCP server. If a DHCP server is found, IP
Address, subnet mask and gateway are auto-
matically obtained. If not found, the device
will check for a DHCP server every 30
seconds. When found, the IP Address, subnet
mask and gateway are automatically ob-
tained.
IP Address using Static method: Upon power Figure 163: Wired Ethernet Screen
up of the system, the device will immediately • The alarm IP Address defaults to Static
begin using the fixed IP configuration. method. To select the DHCP method,
touch the DHCP tab.
CAUTION: • To change the IP Address, touch the IP Ad-
Only facility information systems personnel dress tab. Using the keyboard, enter a new
should set up the network interface. If the IP address.
set up is done improperly, equipment may • To change the Subnet Mask, touch the
not perform properly. Subnet Mask tab. Using the keyboard,
enter a new Subnet Mask address.
• To change the Default Gateway, touch the
CAUTION:
Default Gateway tab. Using the keyboard,
The information systems personnel should enter a new Default Gateway.
be notified before changing any of the net- • To change the DSN Server Address, touch
work settings. Changing the settings could the DSN Server Address tab. Using the key-
keep the equipment from working properly. board, enter a new DSN Server Address.
4107 9016 58.02 4-40