TotalAlert Infinity™ NFPA Medical Gas Notification System



           4.7.5 “Mirroring” Two Alarm Panels                   •  Touch Wired Ethernet tab to access the
                                                                   Wired Ethernet control screen (Section
           Each alarm panel has an Ethernet port on the            4.7.3).
           back of the front panel assembly. By con-            •  Configuration of the Wired Ethernet set-
           necting the alarm to the hospital network,              tings to be performed by facility’s IT spe-
           via Ethernet, or directly cabling it to another         cialist. Settings for the IP Address, Subnet
           alarm panel, signals and gas pressures can              Mask, Default Gateway, and DNS Server
           be pulled from one panel and displayed on               Address to conform to facility’s network.
           another panel                                        •  Touch the Save button to close and return
                                                                   to the MISCELLANEOUS Screen.
           An category 5 CROSS OVER cable is required           •  Exit the MISCELLANEOUS screen and re-
           for connecting alarm panels directly to each            turn to the Configuration screen.
           other. Otherwise, standard cabling can be
           used with a facility’s Ethernet hub or switch
           to connect panels.

           Setup of alarm to be “mirrored”

           Start on the Configuration screen.
           •  Enter Security Password








                                                                       Figure 171: Configuration Screen

                                                               •  Touch the REMOTE SIGNALS tab.








                    Figure 169: Security Password
           •  Touch the MISCELLANEOUS tab located
              on the setup screen





                                                                      Figure 172: Remote Signals Screen

                                                               Each alarm panel can mirror signals and pres-
                                                               sures from up to 8 remote alarm panels.

                                                               •  Touch the E01 tab.
                                                               •  Enter the Remote Ethernet IP address of
                                                                   the alarm panel to be mirrored.
                                                               •  Touch the save button.
              Figure 170: Wired Ethernet Control Screen

           4107 9016 58.02                                 4-42