TotalAlert Infinity™ NFPA Medical Gas Notification System
4.7.5 “Mirroring” Two Alarm Panels • Touch Wired Ethernet tab to access the
Wired Ethernet control screen (Section
Each alarm panel has an Ethernet port on the 4.7.3).
back of the front panel assembly. By con- • Configuration of the Wired Ethernet set-
necting the alarm to the hospital network, tings to be performed by facility’s IT spe-
via Ethernet, or directly cabling it to another cialist. Settings for the IP Address, Subnet
alarm panel, signals and gas pressures can Mask, Default Gateway, and DNS Server
be pulled from one panel and displayed on Address to conform to facility’s network.
another panel • Touch the Save button to close and return
to the MISCELLANEOUS Screen.
An category 5 CROSS OVER cable is required • Exit the MISCELLANEOUS screen and re-
for connecting alarm panels directly to each turn to the Configuration screen.
other. Otherwise, standard cabling can be
used with a facility’s Ethernet hub or switch
to connect panels.
Setup of alarm to be “mirrored”
Start on the Configuration screen.
• Enter Security Password
Figure 171: Configuration Screen
• Touch the REMOTE SIGNALS tab.
Figure 169: Security Password
• Touch the MISCELLANEOUS tab located
on the setup screen
Figure 172: Remote Signals Screen
Each alarm panel can mirror signals and pres-
sures from up to 8 remote alarm panels.
• Touch the E01 tab.
• Enter the Remote Ethernet IP address of
the alarm panel to be mirrored.
• Touch the save button.
Figure 170: Wired Ethernet Control Screen
4107 9016 58.02 4-42