TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                 •  Touch the ETHERNET tab that corresponds
                                                                    to the alarm panel providing the gas pres-
                                                                    sure to be mirrored.














                 Figure 173: Remote Ethernet IP Address
            On the alarm panel that will display the
            mirrored signal or pressure:
            Gas Badge                                                      Figure 176: Ethernet Screen
            •  Go to Configuration screen


















                                                                        Figure 177: Configure Area Screen
                    Figure 174: Configuration Screen
            •  Add new badge to be mirrored by touch-            •  At the CONFIGURE (AREA) screen (B60
               ing the [+] key.                                     Gas Input), touch the gas to be mirrored.
            •  Touch the badge number to be config-              •  Touch the save button.
               ured.                                             •  Follow the steps for configuring gas badg-
            •  On the CONFIGURE (AREA) screen, touch                es listed in Section 4.3.
               A=None/Blank tab to set up badge.



















                   Figure 175: Configure Area Screen

                                                             4-43                               4107 9016 58.02