TotalAlert Infinity™ NFPA Medical Gas Notification System



            4.7.4 BACnet Set-Up


            The TotalAlert Infinity™ alarms employ the
            TCP/IP family of protocols for communica-
            tion and each device has a unique IP address.
            The alarm utilizes a built-in Ethernet connec-
            tion for BACnet transmission of signals to a
            Building Automation System. The TotalAlert
            Infinity™ alarm panel conforms to the BACnet
            protocol and includes a Protocol Implemen-
            tation Conformance Statement (PICS).                            Figure 166: BACnet Screen

            •  Touch the MISCELLANEOUS tab located               •  Save the BACnet device name by touching
               on the configuration screen.                         the save button located in the top right
                                                                    corner of the screen.

















                    Figure 164: Configuration Screen

            •  Touch the BACNET tab to access the BAC-                  Figure 167: BACnet Device Name
               NET control screen.                               •  To add the ID, touch the DEVICE ID tab
                                                                    and enter the ID number provided by the
                                                                    Building Management System administra-
                                                                    tor.














                    Figure 165: Miscellaneous Screen

            •  To allow BACnet communication, touch
               the BACnet ON tab.
            •  To add the name, touch the DEVICE                          Figure 168: BACnet Device ID
               NAME tab and enter the name through               •  Save settings.
               the on screen keyboard.                           •  Touch the X in the top right corner to exit
                                                                    out of the BACnet setup screen.