TotalAlert Infinity™ NFPA Medical Gas Notification System
F = Low Alarm Settings Tab
You can remove the low alarm set point or
customize the low alarm set point.
To remove the low alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SET-
TINGS CHANGE screen.
To customize the low alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
Figure 151: Source Location 3. Touch the Save button.
E = [Units of Measure] Tab
To set up the 4-20mA device:
1. Touch the tab for “4MA=” to set the low
measurement point of the device (zero
for example) and press the save button to
exit.
2. Touch the tab for “20MA=” to set the high
measurement point of the device (100
for example) and press the save button to
exit.
3. Touch the tab for “Units=” to set the
method of measurement (percent for Figure 153: Low Alarm Settings
example), type in the measurement type
and press the save button to exit. G = High Alarm Settings Tab
4. Touch the X button in the upper right to You can remove the high alarm set point or
close the screen. customize the high alarm set point.
To remove the high alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SET-
TINGS CHANGE screen.
To customize the high alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
3. Touch the Save button.
Figure 152: Units of Measure Tab