TotalAlert Infinity™ NFPA Medical Gas Notification System
4.7 Additional Components After configuring the Alarm Type, the B50
board must be initialized. Select “Initialize
4.7.1 4-20 mA Devices Modules” on the Primary Settings Screen.
Press the “Scan” button to make sure the B50
To configure a 4-20mA device for monitoring, board is set up, as it should turn green. Once
start on the CONFIGURATION screen. Users the B50 board is green, press the “Save” icon
are always prompted to enter the password to exit the screen. Exit the “Primary Settings”
when accessing this screen. screen.
Type the password, and touch the Enter but-
ton.
The CONFIGURATION screen displays.
Figure 146: initialize Modules screen
To configure the first badge to custom set-
Figure 144: Configuration screen tings, touch the corresponding number but-
ton in the numeric display that represents
The TotalAlert Infinity™ alarm is designed to the badge to be configured. If a badge needs
allow configuration of four gas badges using to be added, press the “+” for additional
the B50 board to monitor 4-20mA devices. badges.
The alarm must be set as an Area panel or
Combination panel. See Alarm Type Tab in Eight tabs are used for configuration of a
Section 4.3.2. custom badge. Tabs are listed in alphabetical
order.
Figure 145: Primary Settings screen
Figure 147: Configure Area screen
4-35 4107 9016 58.02