TotalAlert Infinity™ NFPA Medical Gas Notification System
• Touch the 01 tab. Enter the instruction. • Touch the X button to the left of the ap-
Touch the [<ÁÑ] to toggle to symbols. propriate instruction tab. When your se-
Touch the [1AZ] key to toggle back to let- lection appears checked and highlighted,
ters and numbers. touch the Save button.
• This specific instruction (rather than the
general instructions) will now display on
the ACTIVE ALARM screen when the alarm
is in fault status. It will also display on the
ALARM POINTS screen.
Figure 141: Optional Alarm Message
• Touch the Save button to close.
• Repeat to add additional instructions.
• Touch the X button to close and return to
the ALARM MESSAGES screen.
Figure 143: Alarm Message: High
• Touch the X button to close and return to
the ACTIVE ALARM OPT screen. • Changing the Alarm Default Screen
• To tie an optional alarm message, or • The TotalAlert Infinity™ Area Alarm switch-
specific instruction, to a set point, go to es to the ACTIVE ALARM screen if the
the LOW ALARM SETTINGS screen or the alarm goes into fault status.
HIGH ALARM SETTINGS screen for that gas • To configure the alarm so that the MAIN
• Touch the OPTION MSG tab screen appears in the event of a fault,
touch the MAIN screen tab.
• If the alarm is already displaying the MAIN
screen when it goes into fault status, the
toolbar will switch from Normal to Icon
and the high or low pressure indicator on
the gas badge will flash for the gas pres-
sure out of range.
Figure 142: Low Alarm Settings
4107 9016 58.02 4-34