B)  The Bone Mill stainless steel parts:
                          1.  After use, separate the stainless steel parts (including the blade if needed for reuse) from
                              other contaminated instruments.
                          2.  Place the stainless steel parts in clean water (saline solution is not recommended).
                          3.  Soak/clean the soiled stainless steel parts for 5 minutes in ultrasonic bath with germicidal
                              solution  (i.e.,  3%  Meritz  plus  germicidal  solution)  at  room  temperature;  alternatively,
                              submerge the stainless steel parts in a proper enzymatic detergent (i.e., 1% Metrex dual
                              enzymatic detergent) for 5 minutes at room temperature.
                          4.  Remove all traces of blood and other contaminants with scrub and round nylon brushes
                              using the proper enzymatic detergent. Brush each part for one minute.
                          5.  Rinse with running water.
                          6.  Visually inspect all areas of the stainless steel parts for any remaining traces of
                              contaminants.
                          7.  If needed, repeat all previous steps until properly cleaned.
                          8.  Dry the parts.
               13.  Sterilization
                     Do not sterilize the Battery-power-pack or the Battery Charger
                          a)  Sterilization Inner Frame, HR500-IF
                              The HR500-IF inner frame (10w x 9 ½d x 7h inches) is designed to properly hold and
                              contain the HR500 Motor frame, the Sterile Shield, and all of the HR500 Bone Mill
                              components. It fits inside a standard mid-size, 8 inches height sterilization container such
                              as Wagner WAG1V20ASA, or Sklar Product #10-4668. The components remain trapped
                              in their location when the container is tilted up to 45 degrees in any directions.






























                                                      Sterilization Inner frame, HR500-IF

               22