Don Carson was able to recently help

             his customer experiencing condensation


               on their incubator door using LogMeIn

               Rescue Lens.  Using the AR session,

                Don discovered that the temperature


              probe was inserted into the sample port

                   tubing and hanging down into the

            chamber.  This would prevent appropriate


             air exchange and have contributed to the

                         condensation on the doors.