Configuring a Monitor
Press the ‘Acquire Data’ button. The application will connect to the Guard and get a
fresh set of Flux data. Once the data is collected, the graph will update:
• If the graph reports a measurement status of ‘OK’, then the gain is correct.
• If the graph reports a signal amplitude problem, adjust the gain and then acquire a
fresh set of data. Keep adjusting the gain until the graph reports a measurement
status of 'OK'.
For Flux-R data, also check the analysis; if the leading peaks were not properly
identified, then there may be something wrong with the asset configuration; perhaps the
number of coils is not correct.
Figure 7-25 Sample Successful Flux-R Analysis