Page 20 - type de document
P. 20

EOS          90395-A-PHY-0002-Column feet assembly & setting




                          4)  The remaining quadrant will not play a role in the adjustment and it will be adjusted so

                             that it makes no contact with the leveling plate. Now, on the chosen “no contact”

                             quadrant, turn its leveling bolt CCW by a few turns.

                          5)  There should only be 3 quadrants touching the plate now.

                          6)  Apply even pressure with one hand at the center of the leveling plate while adjusting

                             the levelling bolts with the other hand.

                          7)  With your eyes on the bubble levels, carefully adjust one bubble, in one direction at

                             the time. We identify each separate direction here as “X” and “Y”. You choose which

                             direction to start with.

                          8)  Once the 3 points are leveled, carefully proceed to raise the fourth quadrant (No

                             contact) by turning its bolt CW until it contacts the leveling plate but does not affect the

                             previous adjustment.


































                   d.  Apply the same technique to the other side
   15   16   17   18   19   20   21   22   23   24   25