Page 47 - GuardII+ Series 4208 Platform User Manual
P. 47
User Instructions
Figure 6-10 Select Data to Import
If there are multiple sets of data in the IRIS directory, you can choose to import any number
of them at once. If one or more of the data sets comes from a monitor that is not already in
the database, IAM will be able to add it automatically if desired.
6.5. Push Buttons
Push buttons are located at the top left inside the enclosure and accessible after opening
the front panel. There are four buttons, although only two of them are used for monitor
operation.
Figure 6-11 GuardII+ Internal Buttons
6.5.1. Alert Button
The Alert Button has three functions:
• Pressing the button will clear the monitor Alert state. This is the most common use of
the Alert button.
• Pressing and holding the button for 5 to 10 seconds will set the Alert into Alert Test
mode, and the Alert Relay will perform its programmed behavior. For example, if the
relay is programmed for Momentary Normally-Open operation, holding the button for
6 seconds will force the relay closed for 4 seconds.
• Pressing and holding the button for more than 10 s will set the Alert into Forced Alert
mode, and the Alert Relay is set to its active state until it is reset. For example, if the
41