Oil-Less Claw Medical Vacuum
Appendix A: TotalAlert Embedded Control System
• If the connecting board is incompatible with position to the Manual position. See Figure A.18.
the display board (for example, the boards Moving to this position forces the pump to operate
are not properly connected), only the bottom against the backup vacuum switch, starting and
version button appears with a red button stopping according to the switch settings.
background.
A.3.4 3.5” Main Screen
3.5” Unit screen with Manual Override selected
Figure A.17 3.5” Main Screen
The main (default) screen (Figure A.17) shows
the operation mode of the unit along with its
automatic versus manual mode setting:
• Automatic (Blue = Standby, Grey = Unselected,
Green = Unit Running)
• Off (Red = Selected, Grey = Unselected)
• Manual (Blue = Standby – Backup Vacuum
Switch is not closed, Grey = Unselected, Green
= Unit Running)
A.3.5 3.5” Main Screen: Manual Override
During the system startup, the manual override, Manual Override
located on the printed circuit board (PCB4), Switch
switch is utilized to ensure the pump is in the Off O - On Manual
position. The manual override switch on PCB4 is a
safety measure as well, for emergency situations to X - Off
ensure the pump unit produces medical vacuum. A - Automatic
In the event of an emergency and the control PCB4 with Manual Override switch On
system is not operating eff ectively, the manual
override switch can be moved from the Automatic Figure A.18 Manual Override
A-9 4107 9000 95.04