Page 304 - Womens Pavilion
P. 304
Oil-Less Claw Medical Vacuum
Appendix A: TotalAlert Embedded Control System
A.2.10 5.7” History Screen A.2.11 5.7” Settings Screen
Figure A.12 History Screen Figure A.13 Settings Screen
The history screen (Figure A.12) shows all of The settings screen (Figure A.13) allows the
the system event history excluding service selection of various sub screens that pertain to
maintenance history. system confi guration data. All value/adjustment
changes are password protected.
• Most recent events are shown on the fi rst
(default) page and are shown in descending • Vacuum – Allows the adjustment of system
date/time order. Located at the top right are vacuum operating range.
the page number and the total number of
pages . The maximum number of pages is 200 • Audible – Allows the horn to re-initiate if an
(1000 events divided by 5 events per page). alarm or shutdown event has not cleared, but
the horn was silenced. The allowable values
• Page up and Page down arrows change the for re-initiate time are never, 15 min, 30 min, 1
view 5 events up or down at a time and are hour, 8 hours, and 24 hours.
only visible if subsequent event information is
available. The Home button changes the view • Language – Allows the choice of pre-defi ned
to the top or most recent events and the End display language.
button changes to the bottom or the oldest
events. • Units – Allows the display units to be displayed
and changed.
• A maximum of 1000 most recent events are
stored for the system. • Model – Displays the pertinent system model
information.
• Date/Time – Allows the date/time to be
displayed and changed.
• Network – Allows the network (IP address, etc)
to be displayed and changed.
• General – Allows the miscellaneous general
information to be displayed and changed.
A-7 4107 9000 95.04