Page 526 - Womens Pavilion
P. 526

TotalAlert Infinity™ NFPA Medical Gas Notification System




             CAUTION: Ensure that the pressure on the            5.3.3 Install New Power Supply
             pipeline in the alarm area has been released        1.  Mount the new Power Supply (B05) to the
             before removing any components from the                adapter bracket using the nuts provided
             pipeline connection.                                   (Figures 218 - 221).
                                                                 2.  Locate the Ground cable attached to the
            3.  Remove any existing fittings or adapters
               from the gas tube.                                   power supply and connect the ring ter-
            4.  Install the gas specific DISS demand check          minal end to the rough-in box for chassis
               valve from the supplied tubing package               ground. Use the self-drilling screw pro-
               into the appropriate tube. To prevent                vided.
               gas leaks, wrap Teflon tape a minimum             3.  Remove the 4 corner screws from the
               of three times around the fitting pipe               Power Supply cover and remove the cover.
               threads.                                          4.  Connect the supply power Line and Neu-
            5.  Tighten the adapter until you feel resis-           tral wires to the terminal block.
               tance; then turn the fitting 2 full turns.        5.  Connect the supply power Ground wire
                                                                    to the rough-in box.
            NOTE: If the DISS demand check thread size           6.  Replace the power supply cover and
            is not compatible with the existing tubing              screws.
            connection; use the included fitting kits to
            adapt as needed. Fitting/adapter kit needs
            to be installed upstream from the DISS con-
            nection at the gas sensor.

            6.  Install the gas specific transducer.
            7.  Tighten the DISS nut until you feel resis-
               tance.                                                                               Power
                                                                                                    Supply
            8.  Repeat steps 3-7 until all of the gas spe-
               cific transducers are installed.
            9.  Verify that the gas specific transducers are
               in the appropriate places by referring to
               your records from the disassembly.                                I/O Modules
            10. Return pressure to each gas sensor, one                    Figure 220: TotalAlert Infinity™
               gas at a time and check for leaks. Make                         Single Module Positions
               any adjustments or corrections as needed.

            NOTE: Remote sensors can also be in-
            stalled in compatible BeaconMedæs Zone
            Valve boxes with additional adapter kit
            (4107401625). One required for each sensor.
                                                                                             Power
                                                                                             Supply





                                                                           I/O Modules


                                                                        Figure 221: Double Module Positions
                                                             5-7                                4107 9016 58.02
   521   522   523   524   525   526   527   528   529   530   531