TotalAlert Infinity™ NFPA Medical Gas Notification System
5.2 Retrofit of TotalAlert Area Alarm
2
Panel (4-gas, 8-gas) and MEGA2
Alarms Additional
I/O Module Power
NOTE: The existing Digital Gas Sensors will Supply
be reused for the new alarm.
5.2.1 Remove Components from
Existing Alarm
Mounting
1. Open old alarm panel to be retrofitted Bracket
and ensure that the retrofit kit to be in-
stalled is correct according to the number
of Signal Inputs, Relay outputs, and/or Figure 212: Power Supply, Module Mounting
Gas Sensor signals (See Appendix A). TA2/Mega2 Area 8 Gas
2. Ensure that you have the additional trim a. Install the 2 position metal standoff
kit if retrofitting a TA2/MEGA2, 8-Gas bracket that is included with the addi-
alarm (4107220615). tional trim kit onto the studs the power
3. Test Master alarm signals to validate wir- supply was removed from (Figure 213).
ing and document all signal wires coming b. Mount the new Power Supply (B05)
into the box with alarm type and loca- and the B60 gas input board to the
tions prior to disconnecting the front bracket using the nuts provided with
panel. This information will be required to the modules.
setup the new alarm.
4. Turn off the supply power to the alarm.
CAUTION: Verify that power has been
turned off prior to working on the alarm.
5. Remove and discard the front panel from
the existing alarm. Power
6. Remove the old power supply Supply
5.2.2 Install New Power Supply
Mega Master / Area 4 Gas
a. Install the metal standoff bracket that B60 Mounting
Bracket
is included with the retrofit kit onto the Module
studs the power supply was removed Figure 213: Power Supply, Module Mounting
from.
b. Mount the new Power Supply (B05)
to the bracket using the nuts provided
(Figure 212).
5-3 4107 9016 58.02