TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                 Data Line Layer Options: ISO 8802-3 (“Ether-
                                                                 net”) LAN Character Set Support: ISO 10646
                                                                 (UTF-8) Encoding.
                                                                 1.  Click BACnet Setup to access BACnet
                                                                    Setup page (Figure 201).
                                                                 2.  Enter BACnet Device ID#.
                                                                 3.  Enter BACnet Device Name.
                                                                 4.  Choose to Enable or Disable BACnet Com-
                                                                    munication.
                                                                 5.  Click Submit
             Figure 200: Electronic Notification Setup Page

            BACnet Setup
            The BeaconMedæs TotalAlert Infinity™ NFPA
            control systems utilize the TCP/ IP family of
            protocols for communication and each alarm
            has a unique IP address.
            Each BACnet enabled TotalAlert Infinity™
            alarm conforms to the BACnet protocol.
            •  Protocol Implementation Conformance

            Each BACnet enabled TotalAlert Infinity™                     Figure 201: BACnet Setup Page
            alarm has available a Protocol Implementa-
            tion Conformance Statement (PICS). The PICS          Clear Event Log
            is a written document created by Beacon-             1.  Click Clear Event Log to access the Clear
            Medæs that identifies the particular options            Event Log page (Figure 202).
            specified by BACnet that are implemented in          2.  Click Clear Event Log.
            the TotalAlert Infinity™ alarm.

            •  BACnet Conformance
            BACnet Vendor ID: 543
            Alarm Profile: BACnet Smart Sensor (B-SS)
            BIBB (BACnet Interoperability Building
            Blocks) Supported:
            •  Read Property - B (DS-RP-B)
            •  Dynamic Device Binding - B (DM-DDB-B)
            •  Dynamic Object Binding - B (DM-DOB-B)
            Object Type Support: Analog value, Binary                      Figure 202: Clear Event Log
            value, Device.
                                                                 CAUTION:

                                                                 Be sure to save the event log prior to clear-
                                                                 ing it. It is impossible to recover the event
                                                                 log after it is cleared.