Page 520 - Womens Pavilion
P. 520
TotalAlert Infinity™ NFPA Medical Gas Notification System
5.0 Retrofit 5.1 Retrofit of TotalAlert2 Master
Retrofit Kits Include: Alarm Panel (32 points or 64 points)
NOTE: The existing power supply will be
reused for the new alarm.
5.1.1 Remove Components from Ex-
isting Alarm
1. Open old alarm panel to be retrofitted
and ensure that the retrofit kit to be in-
stalled is correct according to the number
of Signal Inputs, Relay outputs, and/or
Gas Sensor signals (See Appendix A).
Figure 206: TA2/MEGA2/Others 8-Gas Trim Kit
2. Test Master alarm signals to validate wir-
4107 2206 15 Contents ing and document all signal wires coming
Qty Description into the box with alarm type and loca-
1 Trim Plate Retrofit 16x27 tions prior to disconnecting the front
1 Adapter Bracket, Power Supply panel. This information will be required to
2-Position setup the new alarm.
4 Hardware #6-32 x 1/4” long Pan Head Ma- 3. Turn off the supply power to the alarm.
chine screw 4. Disconnect the power cable and the
ribbon cables from the front panel and
remove and discard the front panel.
CAUTION: Verify that power has been
turned off prior to working on the alarm.
Figure 207: TotalAlert Trim Kit
1
4107 2206 16 Contents
Qty Description
1 Trim Plate Retrofit TotalAlert Single
1
1 Trim Plate Retrofit TotalAlert Double
1
1 Adapter Bracket, Power Supply
1 Adapter Bracket, I/O Modules
4 Hardware #6-32 x 1/4” long Pan Head Ma-
chine screw
NOTE: Due to the nature of retrofit installations,
you will have parts left over when finished.
5-1 4107 9016 58.02