Rod Selection:







                                                           Determine the appropriate rod length.  Calipers are
                                                           provided to estimate the rod length if necessary.  (Note
                                                           that the calipers measure rod length on one side and
                                                           can be used to measure cross connector lengths on the
                                                           other)





                                                           The use of pre-cut rods facilitates the surgery by
                                                           minimizing the time necessary to measure and cut a
                                                           long rod to the appropriate length. It is recommended
                                                           to bend the rod to the appropriate lordotic curvature
                                                           (sagittal plane) to lower the overall construct height and
                                                           to improve its biomechanics by reducing the bending
                                                           moment and thus reducing the stresses on the pedicle
                                                           screws. To maximize ease of insertion, the use of pre-
                                                           bent rods are an option, or use straight rods and bend
                                                           the rod such that it passes each screw roughly 90
                                                           degrees from the axis of the screw. Perform rod
                                                           bending outside the operative field using the French
                                                           Rod Bender. This is best accomplished by a series of
                                                           small adjustments which will bend the rod gradually and
                                                           ensure even stress distribution due to bending. Once
                                                           the rod is ready for insertion, hold the rod with the Rod
                                                           Holder and insert the rod into the pedicle screws.





                                                           If the rod is not fully seated in the pedicle screws, then
                                                           use the Rod Pusher to help seat the rods fully in the
                                                           pedicle screws. A Rod Rocker can also be used for
                                                           seating the rod fully in the pedicle screws.