(optional instrument) An additional option for seating
                                                           the rod is the Rod Reducer.  Slide the end of the rod
                                                           reducer over the rod and onto the head of the polyaxial
                                                           screw until it snaps into place on the screw head.
                                                           Repeated operation of the handle will gradually reduce
                                                           the rod into the screw head.  The Long Set Screw
                                                           Inserter can then be used to insert and seat a set screw
                                                           in the screw head.  To remove the Rod Reducer, pull the
                                                           trigger to release the outer sleeve.  After the sleeve is in
                                                           the fully retracted position, the Rod Reducer is removed
                                                           by rotating off of the polyaxial screw head. (note if the
                                                           Rod reducer needs to be removed without a rod and set
                                                           screw installed, a short rod will need to be placed in the
                                                           head of the screw with the rod holder to provide
                                                           counter torque while removing the persuader.)








                                                           Insertion of Set Screws:
                                                           Once all the rods have been seated in the pedicle
                                                           screws, use Set Screw Inserter or Double Ended Set
                                                           Screw Inserter and place set screws onto the tips of the
                                                           inserter and insert into the head of the pedicle screws.
                                                           Do not final tighten until all adjustments are performed.
                                                           Final tighten should only be completed with the torque
                                                           wrench driver below.





                                                           Compression/Distraction:
                                                           After the construct has been assembled, segmental
                                                           compression and distraction may be carried out using
                                                           the Compressor or Distractor.