Lab Equipments Maintenance & Handling Procedure  Suguna Management System   Ver 1.0 / SOP / FQL / P1 - 02  Page  6 of 6


           FPaper and other materials should not be permitted      elevated a minimum of 3 - 4 cm (1½ ") above work
              to  enter  the  exhaust  duct  of  the  hood.  Foreign   surface to allow unrestricted air movement beneath
              objects  can  be  drawn  into  the  duct  work  and   equipment.
              exhaust  fan,  and  will  adversely  affect  the
              performance of the fume hood.                     FWhile personnel are working in the fume hood, the
                                                                   sash  should  be  pulled  down  to  a  level  that  will
           FHoods  are  not  intended  for  the  storage  of       protect the user's face and upper chest. The only
              chemicals or equipment. All excess chemicals and     time the fume hood sash should be fully open is
              equipment should be stored in areas designed for     while setting up equipment for a procedure.
              long term storage.
                                                                FCheck  that  there  is  sufficient  flow  in  the  hood
           FEquipment  and  other  materials  should  not  be      before commencing work.
              stored against the baffle area (slot at back of hood).
              This  baffle  provides  a  means  for  air  movement   FDo not rely on the fume hood exhaust to protect
              through the fume hood. If blocked, the hood will not   you  from  splashes  or  projectiles  --  WEAR
              provide consistent air movement.                     APPROPRIATE  SAFETY  EQUIPMENT  DICTATED
                                                                   BY  TYPE  OF  CHEMICALS  AND  EQUIPMENT
           FLarge equipment placed in fume hoods should be         BEING USED.


























































          Prepared by :                                                             Approved by :