Page 29 - GuardII+ Series 4208 Platform PD User Manual
P. 29

PD Card Operation



                        User Defined Triggers have two configuration parameters:

                        •  Trigger interval
                        •  An optional minimum Qm level
                        The operation of a User Defined Trigger is like that of Time-Based Triggers:

                        •  Every Trigger interval, save a measurement.
                        •  If the minimum Qm level used, then the one of the machines’ Qm results from the
                            measurement must also be above the level.
                        User Defined Triggers are stored in a separate section of the Archive that uses a First In
                     First Out (FIFO) model.  The FIFO contains up to 30 measurement sets; once the Trigger has
                     occurred 30 times, measurements from new Triggers overwrite the oldest measurements in
                     the FIFO.
                        Measurements that are saved because of a User Defined Trigger will appear in PDView
                     with a “UTB” flag.
















                                    Figure 4-9 Example of a User Defined Trigger in PDView

                                     NOTE:
                                     PD is a phenomenon that builds slowly over time.  The other Triggers in
                                     the GuardII+ PD system are designed to capture long-term trend data, and
                                     under  normal  conditions  they  will  provide  enough  information to  detect
                                     when PD is becoming an issue for the machine.

                                     The  User  Defined  Trigger  is  intended  for  short-term  use  or  to  capture
                                    unusual events that the other Triggers would consider outliers.  While the
                                     User  Defined  Trigger  could  be  used  to  acquire  and  download  large
                                     amounts of data, overuse of the Trigger could cause unnecessary inflation
                                     of the database in IAM for no benefit.

                                     NOTE:
                                     When selecting a time interval for the User Defined Trigger, be sure to
                                     take the number of sensors into account; it takes less than 20 minutes to
                                     run a full set of PD measurements when there are 3 inputs, but a little over
                                     an hour to run a full set when there are 12 inputs.  So, setting a 30-minute
                                     time interval when the monitor has 12 inputs running, will not produce a
                                     stable User Defined Trigger result.



                                                              21
                                                                                          www.irispower.com
   24   25   26   27   28   29   30   31   32   33   34