Page 86 - GuardII+ Series 4208 Platform PD User Manual
P. 86

Troubleshooting




                                     NOTE:
                                     Before looking for a sensor synchronization problem, make sure that the
                                     asset being monitored is running at the time; remember that if the asset
                                     isn’t  running,  there  won’t  be  a  synchronization  signal  coming  from  the
                                     sensors.


                        Use the PD Tester to run measurements as described in Section 7.2 on all inputs that
                     report a synchronization failure.

                        •  If  one  or  more  measurements  fail  due  to  lack  of  synchronization,  this  indicates  a
                            problem with the data acquisition module.  Contact Iris Power LP Technical Support
                            for assistance.
                        Using a voltmeter, measure the ac voltage at the end of the input cable from which the
                     monitor  cannot  run  measurements.    The  expected  ac  synchronization  amplitude  can  be
                     predicted based on the line voltage and operating frequency of the asset; common values are
                     shown below.
                                     Line Voltage         Expected Termination Voltage
                                    (kV, phase-to-                   (mVrms)
                                        phase)               50 Hz              60 Hz
                                          3.3               58 ~ 86           69 ~ 103
                                         4.16               74 ~ 110          87 ~ 131
                                          6.6              115 ~ 173          138 ~ 206
                                          11               192 ~ 288          229 ~ 344
                                         13.8              240 ~ 360          288 ~ 432
                                          18               314 ~ 470          376 ~ 564
                                          22               383 ~ 575          459 ~ 689
                                          26               453 ~ 679          543 ~ 815

                        Repeat the same measurement at the termination box.
                        •  If the voltmeter measures the correct voltage and frequency at the termination box, but
                            not at the end of the input cable, replace the coaxial cable between the termination
                            box  and  the  monitor.    Use  an  appropriate  length  of  50  Ω  RG-316  coaxial  cable
                            terminated on both ends with BNC connectors.  After replacing the cable, check to see
                            if the synchronization problem still exists.

                        •  If the voltmeter does not measure the correct voltage and frequency at the termination
                            box or at the end of the input cable, this indicates a problem with the sensor or the
                            cabling  between  the  sensor  and  the  termination;  contact  Iris  Power  LP  Technical
                            Support for assistance.

                                     WARNING:
                                     Under NO circumstances should any troubleshooting be performed on the
                                     termination box while the asset is running.  This would be hazardous.

                        The remaining possible cause for the synchronization problems is electrical noise on the
                     synchronization signal.  While the GuardII+ has built-in filters for eliminating synchronization
                     noise, there may be cases where there is noise that the filter does not remove.  At this point,




                                                              78
                                                                                          www.irispower.com
   81   82   83   84   85   86   87   88   89   90   91