Page 496 - Med Plaza and Cancer Center
P. 496
2/DESCRIPTION
Gas Inlet Hose: Provides quick connection and Ground Strap: The two ground straps are used to
disconnection of area alarm panel from the gas lines. maintain ground continuity between the front panel and
rough-in assembly. Note: Contact is required between
Control/Display P.C. Board: This Circuit Board contains ground strap and rough-in box or an alarm condition will
the microprocessor and other components that control be generated.
the electrical operation of the area alarm.
Relay Interface Connector: Used to connect remote
Spring Clamp Connector, 2 wire: Used to connect alarms or monitoring devices to the Line Pressure Alarm
individual switch gauge(s) to the control/display P.C. Panel.
board. (Max: 6 provided.)
Manifold Assembly(s): This assembly is used to
distribute supply gas to the line pressure and switch
gauges.
Figure 4
Rear View-Front Panel
0178-0135-000 Rev. A 2-3