TotalAlert Infinity™ NFPA Medical Gas Notification System



           3.2  NFPA Area Wiring Diagram


























              MAIN DISCONNECT PROVIDED BY OTHERS FIELD POWER WIRING TO BE COPPER RATED FOR 75°C MINIMUM FIELD SIGNAL WIRING TO BE MINIMUM 22AWG  MULTI-CONDUCTOR CABLE MUST BE TWISTED PAIR TYPE WITH SHIELD  TOTAL CIRCUIT LENGTH NOT TO EXCEED 10,000’ & MAX. 1K OHM RESIS- -- -- -- -- INDICATES FIELD WIRING OUTSIDE OF BOX. SIGNAL INPUTS FROM GAS SENSOR MODULES. RELAY OUTPUT CONTACT RATED 3A / 30VDC MAX. SEE MODULE LABEL TO SET MODULE ADDRESS (POSITIONS 2,3, & 4). VERIFY  SET ADDRESS SWT












            Notes:  1.   2.   3.   TANCE  4.   5.   6.   7.   USING TABLE 1.  8.                       Figure 36: Wiring Diagram for TotalAlert Infinity™ Area Alarm
























                                                                                            FACTORY WIRED   CHASSIS GROUND








                                                                                            ENCLOSURE   GROUND SCREW




           4107 9016 58.02                                  3-2