TotalAlert Infinity™ NFPA Medical Gas Notification System
4.0 Operation 4.1.2 Gas Badge
Every gas monitored by the alarm will have
4.1 Overview a gas badge which will show the following
information:
The TotalAlert Infinity™ Medical Gas Notifica- • Gas being monitored
tion System is a touchscreen gas pressure and • Location where gas is being used
system monitoring alarm created for ease of
use. • Gas pressure value
• Gas pressure status (Low, Normal, or
For convenience, it is shipped from the fac- High)
tory as a configured unit.
Figure 39: Gas Badge
4.1.3 Source Signals
Every source signal monitored by the alarm
Figure 38: User Interface will display the following information once
configured:
4.1.1 Main Screen • Equipment being monitored
The first screen that users see on a configured • Location of the source equipment
unit is the MAIN screen. The MAIN screen is • Signal being monitored
the main view for the alarm and shows the • Source signal status Red (open - in alarm)
pressure status of all of the gases or source or Green (closed - normal)
signals the alarm is monitoring. Each source badge can show the status of up
to 8 signals.
NOTE:
The main screen will appear different de-
pending on whether the alarm panel is con-
figured as an area, master or combination.
An area alarm main screen will show gas
pressure badges, a master alarm main
screen will show source badges, and a com-
biantion alarm main screen will show both
gas badges and source badges.
Figure 40: Source Badge