Page 471 - Womens Pavilion
P. 471

TotalAlert Infinity™ NFPA Medical Gas Notification System


           4.1.7 Display Screens                               Information Screen

           Active Alarm Screen                                 The INFORMATION screen provides pertinent
                                                               information about the alarm—including high
           The ACTIVE ALARM screen provides detailed
           information about pressures or signals that         and low alarm values and firmware details—
           are in alarm condition. If an active alarm          in one central place.
           instruction has been set up, users will also
           see an instruction on how to respond to the
           alarm (see section 4.6 - Customized Instruc-
           tions Set Up).













                                                                         Figure 49:  Information Screen


                                                                NOTE:
                    Figure 47:  Active Alarm Screen             The Wireless (WIFI) tab is only present on
           Configuration Screen                                 the information screen if the default set-
           The CONFIGURATION screen is used to con-             ting has been changed on the WIFI setting
           figure the alarm. The configuration screen is        screen to allow WIFI control (See Section
           only accessible with a password. It is divided       4.8).
           into three parts.                                   Version Tab

           The top left section is the Display section. It     Touch the VERSION tab to view details about
           contains a small representative view of the         the:
           MAIN screen and buttons for configuring the         •  Series, code, and style of the alarm
           MAIN screen.

           The top right section is the Settings section.      •  Firmware version for the main display
                                                                   board and additional boards
           Tabs not accessible are grayed out.
           The bottom section is the Icon toolbar.
           Display                                    Settings















                                                                             Figure 50: Version Tab
                              Icon Toolbar
                   Figure 48:  Configuration Screen
           4107 9016 58.02                                  4-4
   466   467   468   469   470   471   472   473   474   475   476