TotalAlert Infinity™ NFPA Medical Gas Notification System
C = [Gas Name] Tab E = [Units of Measure] Tab
Figure 63: Gas Name Tab
Figure 65: Units of Measure Tab
• Touch the Erase button on the keyboard to
delete the gas name. To change the unit of measure:
• Enter the new gas name. Touch the [<ÁÑ] 1. Touch the tab for the appropriate unit of
to toggle to symbols and extended Latin measure.
Characters. Touch the [1AZ] key to toggle 2. Touch the Save button to close the screen.
back to letters and numbers.
• Touch the Save button F = Low Alarm Settings Tab
D = Area Monitored Tab
Figure 64: Area Monitored Tab Figure 66: Low Alarm Settings Tab
To add locations to be monitored: You can remove the low alarm set point or
1. Touch an unused location tab. customize the low alarm set point.
2. Using the keyboard, enter the name of the To remove the low alarm set point:
location that you would like to monitor 1. Touch the NONE tab.
(e.g., ROOM 1). Enter multiple locations 2. Touch the check on the CONFIRM SET-
as needed (e.g., Room 1, Room 5). TINGS CHANGE screen.
3. A check appears when the location is se- To customize the low alarm set point:
lected, the save button has to be pressed
if the selection is new. 1. Touch the ADJUST tab.
2. Enter the new set point.
To edit the location information, touch the 3. Touch the Save button.
Pencil button.
4-9 4107 9016 58.02