TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                Test Alarm Tab
             NOTE:
                                                                Touch the TEST ALARM tab to run a quick test
             The version screens automatically scroll.
             Touch the blinking up or down arrow in the         of the hardware.
             lower right to stop the automatic scroll and       The audible alarm will sound for one-to-two
             start the manual scroll.                           seconds to test that the horn is working prop-
                                                                erly.
            Touch the X button in the upper right to
            close and return to the INFORMATION screen.         The LEDs on the front of the alarm will il-
                                                                luminate as well to show that they are still
            Alarm Points Tab                                    functioning.
            Touch the ALARM POINTS tab for a quick              Warnings Tab
            view of the alarm points—the low and high
            set points—for each gas badge (area alarm) or       Touch the WARNINGS tab to view any pos-
            source signal configuration (master alarm).         sible active warnings that would include lost
                                                                network communication with manually en-
                                                                rolled devices, lost connection to a connect-
                                                                ed/initialized board, or  when a master signal
                                                                input is wired, but not assigned/programmed
                                                                to an alarm signal.













                      Figure 51: Alarm Points Tab

            This screen provides a complete view of the
            configuration and is a fast way to check items
            such as:                                                        Figure 52:  Warnings Tab
            •  Location where the pressure sensors and
               source signals are physically wired.

            •  Gases Monitored
            •  Areas Monitored

            •  Low and high pressure set points for area
               signals

            •  Source Signals Monitored
            Touch the X button in the upper right to
            close and return to the INFORMATION screen.









                                                             4-5                                4107 9016 58.02