TotalAlert Infinity™ NFPA Medical Gas Notification System
Test Alarm Tab
NOTE:
Touch the TEST ALARM tab to run a quick test
The version screens automatically scroll.
Touch the blinking up or down arrow in the of the hardware.
lower right to stop the automatic scroll and The audible alarm will sound for one-to-two
start the manual scroll. seconds to test that the horn is working prop-
erly.
Touch the X button in the upper right to
close and return to the INFORMATION screen. The LEDs on the front of the alarm will il-
luminate as well to show that they are still
Alarm Points Tab functioning.
Touch the ALARM POINTS tab for a quick Warnings Tab
view of the alarm points—the low and high
set points—for each gas badge (area alarm) or Touch the WARNINGS tab to view any pos-
source signal configuration (master alarm). sible active warnings that would include lost
network communication with manually en-
rolled devices, lost connection to a connect-
ed/initialized board, or when a master signal
input is wired, but not assigned/programmed
to an alarm signal.
Figure 51: Alarm Points Tab
This screen provides a complete view of the
configuration and is a fast way to check items
such as: Figure 52: Warnings Tab
• Location where the pressure sensors and
source signals are physically wired.
• Gases Monitored
• Areas Monitored
• Low and high pressure set points for area
signals
• Source Signals Monitored
Touch the X button in the upper right to
close and return to the INFORMATION screen.